Skip to main content

Table 1 Intra-and inter-day assay precision and accuracy for the determination of 5-HT in RPMI medium 1640

From: A simple, efficient and rapid HPLC–UV method for the detection of 5-HT in RIN-14B cell extract and cell culture medium

ASSAY

Nominal concentration

(μg/mL)

Mean measured

concentration

(μg/mL)

CV

(%)

Accuracy

(%)

Intra‐assay (n = 5)

0.1

0.093

2.45

92.96

0.3

0.305

1.12

101.71

4

4.060

1.25

101.50

7.5

7.493

0.32

99.90

Inter‐assay (n = 15)

0.1

0.092

4.63

92.02

0.3

0.310

3.82

103.28

4

3.981

4.69

99.53

7.5

7.518

0.62

100.24

  1. Intra-and inter-day assay precision and accuracy for the determination of 5-HT in RPMI medium 1640