Skip to main content

Table 7 Parameters of system suitability for the proposed HPLC method for the determination of VAL and AML

From: Micellar eco-friendly HPLC method for simultaneous analysis of ternary combination of aspirin, atorvastatin and ramipril: application to content uniformity testing

Parameter

ASP/ATR

ATR/IS

IS/RAM

Reference value

Retention time, min

1.5

2.3

4.3

 

Capacity factor, K’

0.36

1.09

2.9

0–10

Symmetry factor

0.718

0.829

1.05

~ 1

Resolution, RS

2.67

2.4

2. 2

RS > 2

Selectivity, α

3.0

1.75

1.52

α > 1

Theoretical plates, N

400

690.93

603.75