Skip to main content

Table 3 Intra-day and inter-day precisions evaluation

From: The first facile optical density-dependent approach for the analysis of doxorubicin, an oncogenic agent accompanied with the co-prescribed drug; paclitaxel

Precision’s level

Conc. (µg mL− 1)

Rec%

Er%b

SDb

Intra-day precision

5.0

97.284

1.864

1.916

10.0

99.383

2.410

2.425

15.0

98.348

-1.165

0.284

20.0

100.494

0.835

0.831

 

25.0

98.012

-1.98

0.475

Inter-day precision

5.0

96.296

1.960

2.035

10.0

100.494

1.864

1.855

15.0

97.69

-2.305

0.79

20.0

101.358

0.650

0.642

 

25.0

97.95

-2.049

1.2290

  1. *The value is the mean of three replicates measurements, SD; the standard deviation, and Er%; the relative error