Skip to main content

Table 1 Assay validation sheet of the proposed MUPLC method for the determination of PIO, ALO, GLM, and VLD.

From: Earth-friendly micellar UPLC technique for determination of four hypoglycemic drugs in different pharmaceutical dosage forms and spiked human plasma

Parameter

PIO

ALO

GLM

VLD

Linearity range (µg/ mL)

1–80

0.1–25

0.25-50

0.3–50

Correlation coefficient (r )

0.9999

0.9999

0.9999

0.9999

Slope

5.6087

8.3759

4.3799

11.0076

Intercept

0.2519

0.1278

-0.6364

-0.3432

SD of intercept

0.5700

0.0839

0.0930

0.1814

LOD (µg/ mL)

0.3353

0.0331

0.0701

0.0544

LOQ (µg/mL)

1.0162

0.1002

0.2124

0.1648

Accuracy (R %)

101.24

98.96

99.79

100.19

Precision

    

Repeatability (RSD%) a

Intermediate Precision (RSD%) b

1.424

1.822

0.965

1.068

0.552

0.997

0.987

1.256

  1. a The intra-day relative standard deviation of (5, 10, and 15 µg/mL) of PIO, ALO, GLM, and VLD in triplicate using the proposed method
  2. b The inter-day relative standard deviation of (5, 10, and 15 µg/mL) of PIO, ALO, GLM, and VLD in triplicate using the proposed method