Skip to main content

Table 2 Determination of SER.DMP in laboratory prepared mixtures with its acidic degradation products by the proposed HPLC and TLC procedures

From: Chromatographic reversed HPLC and TLC-densitometry methods for simultaneous determination of serdexmethylphenidate and dexmethylphenidate in presence of their degradation products—with computational assessment

SER.DMP. (HPLC method)

Intact (µg/mL)

Acidic degradates (µg/mL)

Degradate %

Intact found (µg/mL)

Recovery % of Intacta

5

20

80

5.04

100.86

10

15

60

10.13

101.25

12.5

12.5

50

12.61

100.91

15

10

40

14.92

99.45

20

5

20

19.90

99.52

Mean

 

100.40

RSD%

 

0.843

SER.DMP. (TLC densitometric method)

Intact (µg/spot)

Acidic degradates (µg/spot)

Degradate %

Intact found (µg/spot)

Recovery % of Intacta

5

20

80

4.99

99.96

10

15

60

10.07

100.07

12.5

12.5

50

12.51

100.21

15

10

40

14.98

99.95

20

5

20

19.60

99.85

Mean

 

100.01

RSD%

 

0.967

  1. aEach experiment is repeated three times