Skip to main content

Table 3 Intraday and interday accuracy and precision of ribavirin from spiked chicken samples (mean ± SD, n = 6)

From: LC–MS/MS separation and quantitation of ribavirin in chicken and comparison of different mass spectrometric platforms

Analyte

QC

Spiking level (?g/kg)

Intraday (n = 6)

  

Interday (n = 6)

  

Found (μg/kg; mean ± SD)

Accuracy (RE %)

Precision

(RSD %)

Found (μg/kg; mean ± SD)

Accuracy (RE %)

Precision

(RSD %)

Ribavirin

LLOQ QC

0.2

0.1843 ± 0.0070

−7.83

3.80

0.1873 ± 0.0087

−6.38

4.67

 

LQC

0.5

0.4657 ± 0.0181

−6.87

3.88

0.4719 ± 0.0207

−5.62

4.39

 

MQC

15

14.40 ± 0.3372

−3.99

2.34

14.61 ± 0.4009

−2.58

2.74

 

HQC

200

202.8 ± 2.7143

1.39

1.34

201.5 ± 3.2029

2.25

1.10

  1. LLOQ QC, lower limit of quantitation quality control; LQC, low quality control; MQC, medium quality control; HQC, high quality
  2. control. RE, relative error; RSD, relative standard deviation