Table 3 System suitability and validation parameters for the proposed HPLC/UV method

From: Green wastewater treatment of repurposed COVID-19 therapy (levofloxacin) using synthesized magnetite pectin nanoparticles, comparison with mesoporous silica nanoparticles

Parameters

HPLC

Reference values*

Linearity (µg/mL)

10–100

 

Correlation coefficient (r)

0.9997

Slope

23.784

Intercept

22.649

Standard deviation of residuals from line

21.325

LOD (µg/mL)

2.959

LOQ (µg/mL)

8.966

Accuracy (Recovery % ± SD)a

101.03 ± 0.86

Precision (RSD)

Intradayb

99.36 ± 0.97

Interdayc

100.11 ± 1.33

tR, min

3.916 ± 0.05

 

Tailing factor (T)

0.78

T ≤ 2, T = 1 for symmetric peak

Capacity factor (Kʹ)

1.42

Kʹ = 1–10 acceptable

Plates number (N)

6255

N > 2000

Height equivalent to theoretical plate (HETP; cm plate−1)

0.03

The smaller the value, the higher the column efficiency

  1. aAverage of three blind concentrations analyzed in triplicate
  2. bAverage of three concentrations (25, 50, 75 µg/mL) analyzed in triplicate on the same day
  3. cAverage of three concentrations (25, 50, 75 µg/mL) analyzed in triplicate on three successive days
  4. * according to USP