Skip to main content

Table 2 Parameters required for system suitability tests of the proposed HPTLC method

From: Greenness assessment of two chromatographic methods developed for the determination of Mupirocin in two binary mixtures along with its impurity

Parameters

Mixture (1)

Mixture (2)

Reference values

[24]

Pseud-D

MUP

FIC

FLU

Pseud-D

MUP

MF

Retardation factor (Rf )

0.15

0.42

0.55

0.74

0.14

0.39

0.68

 

Resolution (Rs)

2.18

1.52

3.8

1.8

2.32

Rs > 1.5

Tailing factor (T)

0.83

0.95

0.83

0.84

0.8

1

0.83

 

Retention factor (k’) a

5.66

1.38

0.818

0.35

6.14

1.56

0.47

1–5

Selectivity factor (α)

4.10

1.68

2.33

3.93

3.32

α > 1

Column efficiency (N) b

1024

711

4096

3505

1600

256

4096

 

Height equivalent to theoretical plate (HETP) (mm)

0.0007

0.001

0.0001

0.0002

0.0005

0.0031

0.0002

The smaller the value, the higher the column efficiency

  1. aα = k2/k1, where k is the capacity factor: k = (1 − Rf)/Rf.bRs = [2 (Rf2 − Rf1)]/ (W1 + W2), where Rf is retardation factor and W is peak width at 5% from the baseline of the peak height
  2. CN = 16 (z/w) 2, where z is the migration length of the spot and w is the spot width