Skip to main content

Table 3 Analysis of LDC and MIC in their dosage form by the proposed TLC and RP-HPLC methods with the application of standard addition technique for RP-HPLC method

From: Green chromatographic methods for determination of co-formulated lidocaine hydrochloride and miconazole nitrate along with an endocrine disruptor preservative and potential impurity

Drug

TLC-Densitometric

RP-HPLC

Claimed Concentration

µg/ band

Recovery %a

Claimed concentration (µg mL−1)

Recovery%b

Added concentration (µg mL−1)

Recovery%

MIC

2.50

98.80 ± 0.4

50.00

100.57 ± 1.301

15.00

98.20

30.00

100.13

Mean ± SD

45.00

101.84

 

Mean ± SD

100.06 ± 1.821

LDC

Claimed Concentration

µg/ band

Recovery %a

Claimed concentration (µg mL−1)

Recovery%b

Added concentration (µg mL−1)

Recovery%

0.66

99.49 ± 0.878

13.20

99.91 ± 1.334

8.00

99.50

16.00

101.75

Mean ± SD

24.00

99.25

 

Mean ± SD

100.17 ± 1.349

  1. aaverage of three determinations