Skip to main content

Table 1 Regression and validation parameters of the proposed HPLC and HPTLC methods for determination of MIC, NYS, and MET

From: Ecofriendly single-step HPLC and TLC methods for concurrent analysis of ternary antifungal mixture in their pharmaceutical products

Parameters

HPLC

HPTLC

MIC

NYS

MET

MIC

NYS

MET

Linearity

5–50 μg/mL

4–50 μg/mL

4–40 μg/mL

0.4–2 μg/band

0.4–2.2 μg/band

0.4–2 μg/band

Slope

13.9680

4.3515

28.2340

0.4762

0.8460

1.102

Intercept

− 0.9486

193.2800

3.8488

0.2186

− 0.2224

0.4583

Correlation coefficient

0.9999

0.9999

0.9999

0.9999

0.9998

0.9999

Accuracy (%)

100.41

100.42

99.19

99.53

100.28

100.00

Repeatability (RSD%)a

0.17

0.62

0.61

0.80

0.49

0.77

Intermediate Precision (RSD%)b

0.71

0.82

0.66

1.33

0.67

1.69

LODc

1.19 μg/mL

0.95 μg/mL

0.80 μg/mL

0.11 μg/band

0.09 μg/band

0.07 μg/band

LOQc

3.62 μg/mL

2.87 μg/mL

2.43 μg/mL

0.34 μg/band

0.27 μg/band

0.22 μg/band

  1. aThe intraday precision (n = 3), average of three different concentrations repeated three times within day
  2. bThe interday precision (n = 3), average of three different concentrations repeated three times in three successive days
  3. cLimit of detection and quantitation are determined via calculations: LOD = (SD of the response/slope) × 3.3; LOQ = (SD of the response/slope) × 10