Skip to main content

Table 1 Mobile phase compositions used in method optimization process

From: Stability indicating high performance thin layer chromatography method development and validation for quantitative determination of tetracycline hydrochloride in tetracycline hydrochloride active pharmaceutical ingredient (API) and its dosage forms

S.No.

Mobile phase composition

Proportions (v/v)

Observations

1

Acetone: EDTA: methanol: 1% ammonia solution

65:5:10:10

Poor resolution of peaks

2

Acetone: EDTA: methanol: 1% ammonia solution

65:5:10:15

Resolution problem between peaks

3

Chloroform: acetonitrile: methanol: 1% ammonia solution

11:49:25:15

Poor resolution between peaks

4

Chloroform: dimethylformamide: acetone: 1% ammonia solution

20:20:40:30

Longer development time, high noise

5

Ethyl acetate, acetone, methanol, and 1% ammonia solutiona,b

4.4:19.6:10:6

Sharp peak with good resolution at Rf 0.27 \(\pm\) 0.02

6

Ethyl acetate: acetonitrile: methanol: 1% ammonia solution

4.4:19.6:10:6

impurity peak overlap on analyte peak

7

Ethyl acetate: dimethylformamide: acetone: 1% ammonia solution

10:22:50:18

Peak resolution problem

8

Water: methanol: acetone: dichloromethane

10:70:20:60

Longer development time

  1. aSelected mobile phase composition
  2. bRf for unknown impurity (0.005), ETC (0.123), TC-HCl (0.273), 4-EATC (0.352) and ATC (5, 0.492) with optimized mobile phase; resolution solution is the mixture of analyte and impurities